EasyManua.ls Logo

Agilent Technologies 53131A - To Retrofit the Rear Terminals (Option 060)

Agilent Technologies 53131A
252 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 Retrofitting Options
To Retrofit the Rear Terminals (Option 060)
4-14 Assembly-Level Service Guide
4
To Retrofit the Rear Terminals (Option 060)
Option 060 Rear Terminals Parts
Item Agilent Part No. Qty.
• Coaxial cable assembly (W2) 8120-6150 2
• BNC washer (H6) 2190-0068 2
• BNC nut (H3) 2940-0256 2
1 Remove the cover.
See the section titled “To Remove the Cover” in Chapter 3 of this guide.
2 For retrofitting the Agilent 53131A or Agilent 53132A that do NOT contain the
Optional Channel 3 (Option 030), push out INPUTS plugs 1 and 2 (located on
rear of chassis) as shown in Figure 4-3A.
For retrofitting the Agilent 53131A or Agilent 53132A that contain the Optional
Channel 3 (Option 030), push out INPUTS plugs 1, 2, and 3 (located on rear of
chassis) as shown in Figure 4-3A.
For retrofitting Agilent 53181As that do NOT contain the optional Channel 2
(Option 015/030), push out INPUTS plug 1 (located on rear of chassis) as shown
in Figure 4-3A.
For retrofitting the Agilent 53181As that contain the optional Channel 2, push
out INPUTS plugs 1 and 2 (located on rear of chassis) as shown in Figure 4-3A.
See the following NOTE.
NOTE The Agilent 53131A and 53132A front and rear terminals can be in parallel for inputs
1 and 2. However, the optional Channel 3 input (Option 030) can be configured as
rear terminals only or front terminals only. The optional Channel 3 inputs (Option
050 or 124) can be front terminal only (53131A, 53132A, 53181A).