Chapter 7 Agilent 53131A/132A Specifications
Measurement Specifications
Assembly-Level Service Guide 7-5
7
The following graphs may also be used to compute errors for Period Measurements. To find the Period error (∆P), calculate the frequency of the
input signal ( ) and find the frequency error (∆F) from the chart. Then, calculate the period error as:
Measurement Specifications
Frequency, Period
Channel 1 and 2 Range: 0.1 Hz to 225 MHz 4.44 ns to 10 s
Channel 3 Range:
Option 030 100 MHz to 3 GHz 0.33 ns to 10 ns
Option 050 200 MHz to 5 GHz 0.2 ns to 5 ns
Option 124 200 MHz to 12.4 GHz 80 ps to 5 ns
(Period 2 or 3 selectable only via the GPIB interface)
For Automatic or External Arming:
(and signals < 100 Hz using Timed Arming)
For Time or Digits Arming:
LSD Displayed:
RMS Resolution:
Agilent 53131A Agilent 53132A
t
res
t
res
typical 650 ps 200 ps
see graphs for worst case resolution performance
For Automatic Arming:
where N = 1 for Ch1 or Ch2 Frequency < 1 MHz
4 for Ch1 or Ch2 Frequency > 1 MHz
128 for Ch3
LSD Displayed:
RMS Resolution:
Agilent 53131A Agilent 53132A
t
res
t
jitter
t
res
t
jitter
typical 500 ps 50 ps 225 ps 3 ps
see graphs for worst case resolution performance
Number of Samples =
Gate Time x Frequency (Frequency < 200 kHz)
Gate Time x 200,000 (Frequency > 200 kHz)
Systematic Uncertainty:
Agilent 53131A Agilent 53132A
t
acc
t
acc
typical 350 ps 100 ps
worst case 1.25 ns 500 ps
Trigger: Default setting is Auto Trigger at 50%
Systematic Uncertainty:
Agilent 53131A Agilent 53132A
t
acc
t
acc
typical 100 ps 10 ps
worst case 300 ps 100 ps
Trigger: Default setting is Auto Trigger at 50%
t
res
Gate Time
----------------------------


Frequency
or
Period
×
t
res
2
2 Trigger Error
2
×()+
Gate Time
-----------------------------------------------------------------------



Frequenc
y
or
Period
×
G
ate Time
N
Frequenc
y
-------------------------
---
=
22 t
res
×
Gate Time Number or Samples×
--------------------------------------------------------------------------------------------
t
jitter
Gate Time
----------------------------
+



Frequency
or
Period
×
4t
res
2
2 Trigger Error
2
×+×
Gate Time Number or Samples×
--------------------------------------------------------------------------------------------
t
jitter
Gate Time
----------------------------
+



x
Frequenc
y
or
Period
Time Base Error±
t
acc
Gate Time
----------------------
±


Frequency
or
Period
× Time Base Error±
t
acc
Gate Time
----------------------
±


Frequency
or
Period
×
F
1
P
-
---
= P∆
F
∆
F
-------


P×=