Chapter 7 Agilent 53131A/132A Specifications
Measurement Specifications (Continued)
7-10 Assembly-Level Service Guide
7
Pulse Width
Measurement is specified over the full signal range of Channel 1. The width of the opposing pulse must be greater than 4 ns (e.g., when
measuring the positive pulse width, the negative pulse width must be greater than 4 ns).
Pulse Selection: Positive or Negative
Trigger: Default setting is Auto Trigger
8
at 50%
Results Range: 5 ns to 10
5
s
LSD:
RMS Resolution:
t
res
Systematic Uncertainty:
± (Time Base Error × Pulse Width) ± Trigger Level Timing Error±1.5 ns Differential Channel Error (Agilent 53131A)
± 900 ps Differential Channel Error (Agilent 53132A)
Rise/Fall Time
Measurement is specified over the full signal ranges of Channel 1. The interval between the end of one edge and start of a similar edge
must be greater than 4 ns. (e.g., when measuring a rising edge, 4 ns must elapse between the 90% point of one rising edge and the
10% point of the next rising edge).
Edge Selection: Positive or Negative
Trigger: Default setting is Auto Trigger
8
at 10% and 90%
Results Range: 5 ns to 10
5
s
LSD:
RMS Resolution:
t
res
Systematic Uncertainty:
± (Time Base Error × Transition Time) ± Trigger Level Timing Error±1.5 ns Differential Channel Error (Agilent 53131A)
± 900 ps Differential Channel Error (Agilent 53132A)
Phase
Measurement is specified over the full signal range of Channels 1 and 2.
Results Range: –180° to +360°
RMS Resolution:
t
res
Systematic Uncertainty: (± Trigger Level Timing Error ± 1.5 ns Differential Channel Error) × Frequency × 360° (Agilent 53131A)
Trigger Level Timing Error ± 900 ps Differential Channel Error) × Frequency × 360° (Agilent 53132A)
8
Restrictions noted on page 7-2 for Auto Trigger apply to the proper operation of these measurements. The Peak Volts measurement is used to
determine the signal amplitude and inaccuracies from this, noted on page 7-11, should be included in calculating theTrigger Level Timing Error.
Measurement Specifications (Continued)
53131A 53132A
500 ps 150 ps
----------------------------------------------------------------------------------------
----
t
res
2
Start Trigger Error
2
Stop Trigger Erro
r
2
++
53131A 53132A
750 ps 300 ps
----------------------------------------------------------------------------------------
----
53131A 53132A
500 ps 150 ps
----------------------------------------------------------------------------------------
----
t
res
2
Start Trigger Error
2
Stop Trigger Erro
r
2
++
53131A 53132A
750 ps 300 ps
----------------------------------------------------------------------------------------
----
t
res
2
2 Trigger Error
2
×()+()1
Phase
360°
-----------------


2
+


× Frequency 360°××
53131A 53132A
750 ps 300 ps
----------------------------------------------------------------------------------------
----