EasyManua.ls Logo

Agilent Technologies 53131A - Page 31

Agilent Technologies 53131A
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1 Performance Tests
Agilent 53131A/132A/181A Operational Verification
Assembly-Level Service Guide 1-7
1
Figure 1-1B. Operational Verification Setup for Agilent 53181A
8 Depending on which Counter you are testing, verify the appropriate reading as
shown below:
For Agilent 53131A and Agilent 53181A, verify a reading of 10.000,000,00 MHz.
For the Agilent 53132A, verify a reading of 10.000,000,000 MHz.
9 For Agilent 53131A/132A, mark Pass or Fail in the Agilent 53131A/132A
Performance Test Record on page 1-41, Test 2.
9’ For Agilent 53181A, mark Pass or Fail in the Agilent 53181A Performance Test
Record on page 1-71, Test 2.
NOTE The remaining steps apply to the Agilent 53131A and Agilent 53132A only since
the Agilent 53181A does not have a standard Channel 2 input. For the
Agilent 53181A, proceed to the following “Run Self Test” section.
10 Press Freq & Ratio key until FREQUENCY 2 is displayed to select Channel 2 for
frequency measurements.
For the Agilent 53131A, verify a reading of 10.000,000,00 MHz.
For the Agilent 53132A, verify a reading of 10.000,000,000 MHz.
11 For Agilent 53131A/132A, mark Pass or Fail in the Agilent 53131A/132A
Performance Test Record on page 1-41, Test 3.
Channel 1
H
P 53181A
C
ounter
10 MHz
Out
(rear panel)
深圳德标仪器 135-1095-0799

Related product manuals