Chapter 1 Performance Tests
Agilent 53131A/132A/181A Operational Verification
Assembly-Level Service Guide 1-11
1
Procedure
1 Set the Pulse Generator to the following:
PERIOD: 20 µs
WIDTH: 5 µs
High: 3.00 V Low : 0.8 V
Trailing/Leading : 1.0 ns
Input Mode: TRIG
2 Connect the equipment as shown in Figure 1-3.
Figure 1-3. External Arm Test Setup (Agilent 53131A/132A Only)
HP 8130A
Pulse Generator
C
ounter
HP 8663A
Synthesized Signal Generator
N-to-BNC
Connect
or
Output
Channel 1
50
Feedthrough
Ext Arm
(rear panel)
Output