Chapter 1 Performance Tests
Test 3: Channels 1 and 2 Frequency Sensitivity (Agilent 53131A/132A
Only)
Assembly-Level Service Guide 1-23
1
100 kHz to 100 MHz Sensitivity for Agilent 53131A/132A
1 Connect the timebase output of the unit with the better timebase to the timebase
reference input of the other unit as shown in Figure 1-7.
For this test, make sure you always use the unit (Agilent 53131A/132A or
Agilent 8663A) that contains the better 10 MHz timebase as the output source as
shown in Figure 1-7.
Figure 1-7. Channels 1 and 2 Frequency Sensitivity Test Setup
H
P 53131A/132A
C
ounter
HP 8663A
Synthesized
Signal Generator
N-to-BNC
Connect
or
Output
Channel 1
50
Feedthrough
Ref. In
(rear panel)
50
Feedthrough
10 MHz
Output
(rear panel)
Output
(rear panel)
H
P 53131A/132A
C
ounter
HP 8663A
Synthesized
Signal Generator
N-to-BNC
Connecto
r
Output
Channel 1
Input
(rear panel)
OR