Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (Agilent
53131A/132A Only)
Assembly-Level Service Guide 1-31
1
100 MHz to 2.5 GHz Sensitivity for Agilent 53131A/132A Option
030 only
1 Set the Agilent 8663B Synthesized Signal Generator to output a 2500 MHz, 40
dBm sine wave, and connect the signal to Channel 3 of the Counter as shown in
Figure 1-9.
Figure 1-9. Option 030/050/124 Channel 3 Frequency Sensitivity and Accuracy Test Setup
2 On the Counter, Press Run key.
P 53131A/132A
ounter
HP 8663A
Synthesized Signal Generator
N-to-BNC
Connector
Output
Channel 3
50
Feedthrough
Ref. In
(rear panel)
Input
(rear panel)
HP 8340B
Sweep Oscillator
Frequency Standar
(switch set to EXT)
Output