EasyManua.ls Logo

Agilent Technologies 53131A - Page 57

Agilent Technologies 53131A
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (Agilent
53131A/132A Only)
Assembly-Level Service Guide 1-33
1
200 MHz to 5.0 GHz Sensitivity for Agilent 53131A/132A (Option
050 only)
1 Connect the 8340B Sweep Oscillator to Channel 3 of the counter.
2 Change the settings of the Sweep Oscillator to 5.0 GHz, 40 dBm.
3 Increase the power level in 1 dB steps until the Counter displays a stable reading
of approximately 5.0 GHz.
The Counter should display 5.0 GHz at 23 dBm.
NOTE Use a power meter and sensor to ensure the accuracy of the power being input.
4 Sweep the frequencies from 200 MHz to 5.0 GHz.
The Counter should read frequencies from 200 MHz to 5.0 GHz at
≤−23 dBm.
5 Record the actual reading in the Performance Test Record (Test 5, Line 3).
6 Disconnect the test setup.
深圳德标仪器 135-1095-0799

Related product manuals