Chapter 1 Performance Tests
Test 7: Peak Volts, Channels 1 and 2 (Agilent 53131A/132A Only)
Assembly-Level Service Guide 1-39
1
Volt Peak 2
1 Remove the signal from Channel 1 of the Counter and connect it to Channel 2.
2 Press Other Meas key until VOLT PEAK 2 is displayed.
The Counter should display -1.00 ±0.12V for the negative peak of the sine wave. This
reading is displayed on the left side of the display.
Record the actual reading in the Performance Test Record (Test 7, Line 2a).
Also, the Counter should display +1.00 ±0.12V for the positive peak of the sine wave.
This reading is displayed on the right side of the display.
Record the actual reading in the Performance Test Record (Test 7, Line 2b).
3 Disconnect the test setup.