Chapter 1 Performance Tests
Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity
1-60 Assembly-Level Service Guide
1
3 Increase the power level in 1 dB steps until the Counter reads 500 MHz.
The Counter should display 500 MHz at 27.0 dBm (20 mVrms).
4 Sweep the frequencies from 100 MHz to 1.5 GHz.
The Counter should read frequencies from 100 MHz to 1.5 GHz
at 27.0dBm (20 mVrms).
5 Record the actual reading in the Performance Test Record (Test 4, Line 1).
100 MHz to 2.7 GHz Sensitivity for Option 030 Only
1 Set the Agilent 8663B Synthesized Signal Generator to output a 500 MHz, 60
dBm sine wave, and connect the signal to Channel 3 of the Counter as shown in
Figure 1-14.
2 On the Counter, Press Run key.
3 Increase the power level in 1 dB steps until the Counter reads 500 MHz.
The Counter should display 500 MHz at 27.0 dBm (20 mVrms).
4 Sweep the frequencies from 100 MHz to 2.7 GHz.
The Counter should read frequencies from 100 MHz to 2.7 GHz
at 27.0 dBm (20 mVrms).
5 Record the actual reading in the Performance Test Record (Test 4, Line 2).