Chapter 1 Performance Tests
Agilent 53181A GPIB Verification Program (Optional)
1-70 Assembly-Level Service Guide
1
Agilent 53181A GPIB Verification Program
(Optional)
The GPIB Verification program exercises the Agilent 53181A through various
operating modes via the its GPIB interface. If the Counter successfully completes all
phases of the verification program, there is a high probability that the GPIB interface
is operating correctly. If the Counter fails the verification program, refer to the
troubleshooting section in Chapter 2, “Service,” of this guide.
The GPIB Verification program may be loaded into the HP 9000 series 200 or 300
Desktop Computer from the 3 1/2-inch disk (Agilent P/N 53181-13501) or the 5 1/4-
inch disk (Agilent P/N 53181-13501).
1 To run the program on the disk, insert the disk into the Desktop Computer, load
the program by typing Load “VER81”, and press RUN key on the computers
keyboard.
2 Follow the instructions displayed in the computers screen.
3 After running this program, record the result (Pass or Fail) on the last page
(page 1-74) of the Agilent 53181A Performance Test Record.