Chapter 1 Performance Tests
Agilent 53181A Performance Test Record (Page 2 of 4)
1-72 Assembly-Level Service Guide
1
Agilent 53181A Performance Test Record
(Page 2 of 4)
Complete Performance Tests
Test Number
Test Description Minimum Actual Reading Maximum
1 Trigger Level:
Channel 1 15 mV
15 mV
1. __________
2. __________
+15 mV
+15 mV
Test 2: Channel 1 Frequency Sensitivity (Agilent 53181A Only)
Standard (Front Panel Terminals) Actual Reading Specification
100 kHz–100 MHz, Channel 1 1. __________ 21 dBm (20 mVrms)
100 MHz–200 MHz, Channel 1 2. __________ 17.5 dBm (30 mVrms)
200 MHz–225 MHz, Channel 1 3. __________ 15.0 dBm (40 mVrms)
Option 060 Rear Terminals Actual Reading Specification
100 kHz–225 MHz, Channel 1 1. __________ 9.5 dBm (75 mVrms)
Test
Number Test Description Minimum
Actual Reading
Maximum
3 Channel 1 Frequency
Accuracy
(Agilent 53181A Only)
1 MHz Test
100 MHz Test
200 MHz Test
225 MHz Test
999,999.999 Hz
99.999,999,93 MHz
199.999,999,87 MHz
224.999,999,85 MHz
1. __________
2. __________
3. __________
4. __________
1, 000, 000.001 Hz
100.000, 000,07 MHz
200.000,000,13 MHz
225.000,000,15 MHz
深圳德标仪器 135-1095-0799