FINAL TRIM SIZE : 7.5 in x 9.0 in
Installing
Step 3.
Set
the
line
voltage
C
A
U
T
I
O
N
Do
not
attempt
to
rotate
the
voltage
selector
cam
while
it
is
installed
in
the
line
module
or
non-repairable
damage
will
result.
The
cam
must
be
completely
removed
from
the
line
module
,
rotated
to
the
proper
position,
and
reinstalled.
6.
Remove
the
voltage
selector
cam
from the
line module
.
7.
Rotate
the
voltage
selector
cam
to the
desired voltage
. When
the
line
module
cover
is
closed,
the
selected
voltage will
be visible
through a
small
window
.
8. Insert
the voltage
selector cam
back
into
the
line
module
.
9. V
erify that
the value
of
the
line-voltage
fuse
in
the
pull
out
draw
is
correct.
The recommended fuse for 100V and 120V operation is a F1.5A, 250V
,
part number 2110-0043. The recommended fuse for 220V and 240V
operation is a F0.75A, 250V
, part number 2110-0063.
10. Close the line-module cover door
.