476 Chapter7
Language Reference
FOFFSET Frequency Offset
Parameters
number real from 0 to 2.9E+9 (8560E/EC)
real from 0 to 6.5E+9 (Agilent 8561E/EC)
real from 0 to 13.2E+9 (Agilent 8562E/EC).
real from 0 to 26.5E+9 (Agilent 8563E/EC).
real from 0 to 40E+9 (Agilent 8564E/EC).
real from 0 to 50E+9 (Agilent 8565E/EC).
UP or DN changes by 10 percent of span.
Preset State
Off
Query Response
Figure 7-117 FOFFSET Query Response
Example
10 INPUT "ENTER DESIRED FREQUENCY OFFSET IN HERTZ",Foffset
20 OUTPUT 718;"FOFFSET ";Foffset;"HZ;"
30 OUTPUT 718;"FOFFSET?;"
40 ENTER 718;Foffset
50 PRINT "THE FREQUENCY OFFSET IS",Foffset,"HZ"
60 END