Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
219
Front-Panel Key and SCPI Command
Reference
4.4.5.4.6 Capture Time
Sets the capturing data time. When in manual mode, the user determines the capture time; otherwise, it is
determined automatically by the measurement.
This key is available only when PSA firmware revision A.11.08 or above, or Option 23A is installed.
Key Path: Meas Setup
Remote Command: [:SENSe]:FMDeviation:TRIGger:SOURce
IMMediate|IF|EXTernal[1]|EXTernal[2]|RFBurst
[:SENSe]:FMDeviation:TRIGger:SOURce?
Default: IMMediate
State Saved: Saved in instrument state.
Range: Free Run (Immediate)|Video (IF Envlp)|RF Burst|Ext Front|Ext Rear
Example: :FMD:TRIG:SOUR IF
:FMD:TRIG:SOUR?
Key Path: Meas Setup
Remote Command: [:SENSe]:FMDeviation:CAPTure[:TIME] <time>
[:SENSe]:FMDeviation:CAPTure?
[:SENSe]:FMDeviation:CAPTure:AUTO 0|1|ON|OFF
[:SENSe]:FMDeviation:CAPTure:AUTO?
Preset/Default: 250 ms
State Saved: Saved in instrument state.
Min: 1 μs
Max: 10 s
Resolution 1 ms
Example: :FMD:CAPT 0.25
:FMD:CAPT?
:FMD:CAPT:AUTO OFF
:FMD:CAPT:AUTO?