4.4.6.1 SPAN/X Scale
The SPAN/X Scale key accesses the menu to set the desired horizon scale and associated settings of waveform.
You must have the Demod Waveform selected under the
Trace/View key.
4.4.6.1.1 Scale/Div
This key is for Scale/Div control.
Table 4-22 PM Deviation Numeric Results Window
Name Corresponding Results
PM Deviation n=1 1st
PM Deviation output
Modulation Rate n=1 3rd
Modulation Distortion n=1 4th
Modulation SINAD n=1 5th
Key Path:
Span X Scale
Remote Command: :DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]
:PDIVision <seconds>
:DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]
:PDIVision?
Unit: Seconds
Preset: 100 μs
State Saved: Saved in instrument state.
Min: 1.0 ns
Max: 1 s
Dependencies/Couplings: See Restriction and Notes
Notes: If the Scale Coupling is set to On, this value is automatically determined by the
measurement result. When you set this value manually, Scale Coupling
automatically changes to Off.
Example: :DISP:PMD:WIND:TRAC:X:PDIV 100us
:DISP:PMD:WIND:TRAC:X:PDIV?

Table of Contents

Other manuals for Agilent Technologies E4443A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4443A and is the answer not in the manual?

Agilent Technologies E4443A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4443A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals