4.4.6.1.2 Ref Value
Allows you to set the display reference.
4.4.6.1.3 Ref Position
Allows you to set the display reference position to either Left, Ctr (center), or Right.
Key Path: Span X Scale
Remote
Command:
:DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]:RLEVel
<seconds>
:DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]:RLEVel?
Unit: Seconds
Preset: 0.0
State Saved: Saved in instrument state.
Min: –10 s
Max: 10 s
Dependencies/
Couplings:
See Restriction and Notes
Notes: If the Scale Coupling is set to On, this value is automatically determined by the measurement
result. When you set this value manually, Scale Coupling automatically changes to Off.
Example: :DISP:PMD:WIND2:TRAC:X:RLEV 0s
:DISP:PMD:WIND2:TRAC:X:RLEV?
Key Path:
Span X Scale
Remote
Command:
:DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]:RPOSition
LEFT|CENTer|RIGHt
:DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]:RPOSition
?
Preset: LEFT
State Saved: Saved in instrument state.
Range: Left | Ctr | Right
Notes: None
Example: :DISP:PMD:WIND:TRAC:X:RPOS 5
:DISP:PMD:WIND:TRAC:X:RPOS?

Table of Contents

Other manuals for Agilent Technologies E4443A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4443A and is the answer not in the manual?

Agilent Technologies E4443A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4443A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals