Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
225
Front-Panel Key and SCPI Command
Reference
4.4.6.1.4 Scale Coupling
Allows you to toggle the scale coupling function between On and Off.
4.4.6.2 AMPLITUDE/Y Scale
The AMPLITUDE/Y Scale key accesses the menu to set the desired vertical scale and associated settings:
Default numbers are independent for each window.
4.4.6.2.1 PM Deviation Numeric Results Window
If Numeric Results Window is chosen, the following menu is available.
4.4.6.2.1.1 Display Mode
If measurement results are to be displayed relative to a reference, enter the value as a ratio reference by selecting RATIO.
Key Path:
Span X Scale
Remote Command: :DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]:COUPle
0|1|OFF|ON
:DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]:COUPle?
Preset: ON
State Saved: Saved in instrument state.
Range: Off|On
Dependencies/
Couplings:
See Restriction and Notes
Notes: Upon pressing the Restart front-panel key or Restart softkey under the Meas Control menu,
the scale coupling function automatically determines the scale per division and reference
values based on the measurement results if this parameter is set to On. When you set a value
to either Scale/Div or Ref Value manually, Scale Coupling automatically changes to Off.
Example: :DISP:PMD:WIND:TRAC:X:COUP OFF
:DISP:PMD:WIND:TRAC:X:COUP?
Key Path:
Amplitude Y Scale
Remote Command: :DISPlay:PMDeviation:MODE NORMal|RATio
:DISPlay:PMDeviation:MODE?
Preset: NORMal
State Saved: Saved in instrument state.
Range: Normal | Ratio
Example: :DISP:PMD:MODE NORM
:DISP:PMD:MODE?