Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
179
Front-Panel Key and SCPI Command
Reference
4.4.2.1.3 Ratio Reference
If RF Power is to be displayed relative to a reference, enter the value as a ratio reference using this key.
4.4.2.1.4 Ratio Mode
The ratio reference and ratio mode function permits measurement result to be scaled to a reference. The
reference may be either a measurement result or a keyboard entry. The Log/Linear key enables measurement
results to be displayed in linear or logarithmic units.
Mode: MRECEIVE
Key Path:
Amplitude/Y Scale
Remote Command: :DISPlay:RFPower:RREFerence <real>
:DISPlay:RFPower:RREFerence?
:DISPlay:RFPower:RREFerence:AUTO 0|1|OFF|ON
:DISPlay:RFPower:RREFerence:AUTO?
Unit: dBm
Preset: 0 dBm
State Saved: Saved in instrument state.
Min: –200 dBm
Max: 50 dBm
Dependencies/Couplings: No
Example: :DISP:POW:RREF –10 dBm
:DISP:POW:RREF?
:DISP:RFP:RREF:AUTO OFF
:DISP:RFP:RREF:AUTO?
Mode: MRECEIVE
Key Path:
Amplitude/Y Scale
Remote Command: :DISPlay:RFPower:RMODe LOG|LINear
:DISPlay:RFPower:RMODe?
Preset: LOG
State Saved: Saved in instrument state.
Range: Log | Linear
Dependencies/Couplings: No
Example: :DISP:RFP:RMOD LOG
:DISP:RFP:RMOD?