Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
200
Front-Panel Key and SCPI Command
Reference
4.4.4.3 Trace / View
Allows you to select the desired measurement view from the following selections:
Numeric Results
Demod Waveform
4.4.4.4 Window Selection
Sets the active window.
Notes: Upon pressing the Restart front-panel key or Restart softkey under the Meas Control
menu, the scale coupling function automatically determines the scale per division and
reference values based on the measurement results if this parameter is set to On. When
you set a value to either Scale/Div or Ref Value manually, Scale Coupling
automatically changes to Off.
Example: :DISP:AMD:WIND:TRAC:Y:COUP OFF
:DISP:AMD:WIND:TRAC:Y:COUP?
Mode: MRECEIVE
Key Path:
View/Display
Remote Command: :DISPlay:FMDeviation:VIEW[SELect] NUMeric|WAVeform
:DISPlay:FMDeviation:VIEW[SELect]?
:DISPlay:AMDepth:VIEW[SELect] NUMeric|WAVeform
:DISPlay:AMDepth:VIEW[SELect]?
Preset: NUMeric
State Saved: Saved in instrument state.
Range: Numeric results | Waveform result
Example: :DISP:FMD:VIEW WAV
:DISP:FMD:VIEW?
:DISP:AMD:VIEW NUM
:DISP:AMD:VIEW?
Mode: MRECEIVE
Remote Command: :DISPlay:WINDow:SELect 1|2
:DISPlay:WINDow:SELect?
Preset: 1
State Saved: No