Page 4 of 5 E5062A-04A
(3) PCA in A2 Receiver Module is changed from E5062-66502 to E5062-66512. (Permanent
solution)
Procedures:
For the detailed Replacement Procedure and Post-Repair Procedures, please see Service Guide of
E5061A/E5062A.
[Replacement Procedure]
Chapter 6: Replacement Procedure -> A2 Receiver Board Replacement
[Post-Repair Procedure]
Chapter 7: Post-Repair Procedures -> Table 7-1 A2 Receiver Board.
- Frequency Reference Adjustment
- Local Output Power Adjustment
- Source Output Power Adjustment
- Receiver IF Range Adjustment
- Receiver Ports Characteristics Adjustment
- Diagnostic Test
- Frequency Accuracy Test
E5062-66502 should be changed to
E5062-66512