    ADMINISTRATIVE INFORMATION  
 
SERVICE NOTE CLASSIFICATION: 
MODIFICATION RECOMMENDED 
ACTION  ON SPECIFIED FAILURE 
CATEGORY:     
STANDARDS 
  LABOR:  0.5 Hours               
LOCATION   
CATEGORY:  ON-SITE 
  SERVICE CENTER 
   
AVAILABILITY:  PRODUCT’S SUPPORT LIFE 
NO CHARGE AVAILABLE UNTIL:  end of service life 
PRODUCT LINE:  WN 
AUTHOR:   MF 
ADDITIONAL INFORMATION: Affected PNA Models: E836xC & E8362/3/4CH85. Service Note History: E8364C-02 released 
Jan 2009. E8364C-02A released Dec 4, 2012 to revise serial numbers. 
 © AGILENT TECHNOLOGIES, INC. 2012 
 PRINTED IN U.S.A.   
  December 5,  2012     
  Rev. 20  Page 1 of 5 
E8364C-02A 
S    E    R    V    I    C    E                N    O    T    E    
 
 
            Supersedes: 
E8364C-02  
 
E8364C PNA Microwave Network Analyzers 
 
Serial Numbers:   
MY00000000-MY48500000, SG00000000-SG48500000, US00000000-US48500000 
 
 
A new touchscreen display inverter shield is available to improve PNA noise 
floor performance. 
 
 
Parts Required:  
P/N  Description  Qty. 
     
N5230-00017  Inverter Shield  1