Service Guide E8364-90026 4-11
PNA Series Microwave Network Analyzers Troubleshooting
E8362B, E8363B, E8364B Power Up Troubleshooting
Isolate the remaining three assemblies:
remove the A15 CPU board
remove the A4 power supply
This leaves only the A14 system motherboard installed. If the measurements are still
incorrect, this is the suspected faulty assembly.
Table 4-2 Power Supply Voltages and Resistances
Measurement
Location
Supply
(Vdc)
Approximate
Resistance (Ω)
A+5.0285
B -15.0 12.7 k
C -5.0 7.2 k
D +9 4.0 k
E +15 3.0 k
F +22 2.8 k
G +22 890
H +32 2.6 k

Table of Contents