4 GenericTest
Parallel capacitance Cp=B/(2*PI*Freq)
Parallel resistance Rp=1/G
Dissipation factor D=G/abs(B)
Series resistance Rs=G/(G^2+B^2)
Reactance X=-B/(G^2+B^2)
Series capacitance Cs=-1/(2*PI*Freq*X)
Impedance Z=sqrt(Rs^2+X^2)
Phase Theta=atan(X/Rs)
[X-Y Plot]
X axis: Time (LINEAR)
Y1 axis: Parallel capacitance Cp (LINEAR)
Y2 axis: Conductance G (LINEAR)
[List Display]
Time Time
Parallel capacitance Cp
Conductance G
Frequency Freq
Series capacitance Cs
Series resistance Rs
Parallel resistance Rp
Dissipation factor D
Reactance X
Impedance Z
Phase Theta
DC bias monitor data DcMon
AC level monitor data AcMon