7 MixedSignal
7.12 Ic-Vc Vb Mismatch[3]: Ic-Vce characteristics mismatch, Vb sweep, 3-
terminal (A.01.20)
[Supported Analyzer]
B1500A
[Description]
Measures the collector current vs collector voltage (Ic-Ice) characteristics of device A, and measures the Ic-Ice
characteristics of device B. After that, calculates the differences between Ic values by using the following
formula, and plots the results.
Delta_Ic=(Icollector_A-Icollector_B)/Icollector_A*100
[Device Under Test]
Bipolar transistor, 3 terminals, 2 ea.
[Device Parameters]
Polarity: NPN (SMUs force the specified value) or PNP (SMUs force the negative specified value).
Le: Emitter length
We: Emitter width
Temp: Temperature
IcMax: Collector current compliance
[Test Parameters]
IntegTime: Integration time
CollectorA: SMU connected to Device A Collector terminal, primary sweep voltage output
CollectorB: SMU connected to Device B Collector terminal, primary sweep voltage output
VcStart: Sweep start voltage for Collector terminal
VcStop: Sweep stop voltage for Collector terminal
VcStep: Sweep step voltage for Collector terminal
BaseA: SMU connected to Device A Base terminal, secondary sweep voltage output
BaseB: SMU connected to Device B Base terminal, secondary sweep voltage output
VbStart: Sweep start voltage for Base terminal
VbStop: Sweep stop voltage for Base terminal
VbStep: Sweep step voltage for Base terminal
Emitter: SMU connected to Emitter terminal, constant voltage output
[Extended Test Parameters]
Ve: Emitter voltage
HoldTime: Hold time
DelayTime: Delay time
BaseMinRng: Minimum range for the base current meausrement
CollectorMinRng: Minimum range for the collector current meausrement
[Device A: Measurement Parameters]
Collector current IcollectorA
Base current IbaseA
[Device A: User Function]
hfe_A=IcollectorA/IbaseA
[Device A: X-Y Plot]
X axis: Collector voltage VcollectorA (LINEAR)
Y1 axis: Collector current IcollectorA (LINEAR)
Agilent EasyEXPERT Application Library Reference, Edition 8