11 Reliability
[Sampling Timing adjustment]
SamplingDelay: Adjust this parameter so as the start timing of the sampling measurement observed in the
oscilloscope becomes the same as the sampling start display (is Stress_time_at_ in the Parameter display).
Note: Recommend to use the default 50 ms.
Stress_T_adj: Time (negative) for adjusting the stress time accuracy which depends on the measurement setup
or a PC speed used for Desktop EasyEXPERT.
Note: Stress_T_adj parameter is adjusted so as the Stree_time_at_ display in the Sampling_Ids graph
becomes closer to the end value of the previous stress time. Default value is tuned for the use on the B1500A
EasyEXPERT and the RecordSamplingData=ON condition.
[Test Output: X-Y plot]
X axis: Cumulative stress time or sampling time
Y-axis: Id
Agilent EasyEXPERT Application Library Reference, Edition 8