13 Sollar Cell
13.1 Solar Cell Cp-AC Level: Solar Cell Cp-Vac characteristics
(A.05.03.2013.0124_2013.02.27.1)
[Supported Analyzer]
B1500A
[Description]
Measures the anode-cathode capacitance (Cp) of the solar cell with sweeping the signal level (Vac), and plots
the Cp-Vac characteristics.
DC bias output is fixed at Vdc. Vac is swept for the peak-to-peak voltage Vpp to change from VppStart to
VppStop in VppStep steps. The maximum and minimum value of the applied voltage should be kept Vdc, Vdc-
Vpp respectively. The CMU performs spot measurement of the parallel capacitance (Cp) and conductance (G)
at each signal level.
[Reference]
IV and CV Characterizations of Solar/Photovoltaic Cells Using the B1500A
http://cp.literature.agilent.com/litweb/pdf/5990-4428EN.pdf
[Device Under Test]
Solar Cell
[Required Modules and Accessories]
Agilent B1520A MFCMU 1 unit
[Device Parameters]
Temp: Temperature
Area: Area of the solar cell
Es: Relative permittivity of semiconductor
[Test Parameters]
IntegTime: Integration time
VppStart: Peak-to-peak voltage (Vpp) start value
VppStop: Vpp stop value
VppStep: Vpp step value
FREQ: Measurement frequency
Anode: CMU connected between Anode and Cathode (CV spot measurement)
Vdc: Anode-Cathode voltage, constant voltage output
[Extended Test Parameters]
HoldTime: Hold time
DelayTime: Delay time
CpMin: Minimum value of Cp axis for X-Y graph
CpMax: Maximum value of Cp axis for X-Y graph
GMin: Minimum value of G axis for X-Y graph
GMax: Maximum value of G axis for X-Y graph
[Measurement Parameters]
Parallel capacitance cp
Conductance g
[User Function]
PI=3.141592653589
d=g/(2*PI*FREQ*cp)