EasyManua.ls Logo

Agilent Technologies EasyEXPERT

Agilent Technologies EasyEXPERT
610 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22 IGBT
22.5 Ic(off)-Vce: IGBT Ic(off)-Vce characteristics (A.04.00)
[Supported Analyzer]
B1505A
[Description]
Measures and plots Collector current vs Collector-Emitter voltage in cutoff region and extracts Collector-
Emitter cutoff current and breakdown voltage.
[Device Under Test]
IGBT, 3 terminals
[Device Parameters]
Polarity: Nch (SMU forces the specified value) or Pch (SMU forces the negative specified value).
Temp: Temperature
[Test Parameters]
Memo: Memorandum
IntegTime: Integration time
Collector: SMU connected to Collector terminal, primary sweep voltage output
Vc@Ices: Collector voltage to decide Collector-Emitter cutoff current Ices
Ic@BVces: Collector current to decide Collector-Emitter breakdown voltage BVces
VcStart: Sweep start voltage for Collector terminal
VcStop: Sweep stop voltage for Collector terminal
VcStep: Sweep step voltage for Collector terminal
Gate: SMU connected to Gate terminal, constant voltage output
Vg: Gate voltage
Emitter: GNDU connected to Emitter terminal
IcLimit: Collector current compliance
[Extended Test Parameters]
IgLimit: Gate current compliance
HoldTime: Hold time
DelayTime: Delay time
IcZero: Y axis (Icollector) minimum value
IcMinRange: Minimum range for the collector current measurement
IgMinRange: Minimum range for the gate current measurement
[Measurement Parameters]
Collector current Icollector
Gate current Igate
[User Function]
Ta: Temperature Ta=Temp
[Analysis Function]
BVces=@MX (X coordinate of Marker)
Ices=@L1Y (Y intercept of Line1)
[Auto Analysis]
Marker: Icollector=Ic@BVces
Line1: Vcollector=Vc@Ices
[X-Y Plot]
Agilent EasyEXPERT Application Library Reference, Edition 8
22-11

Related product manuals