23 Interconnection
23.1 Residual R: R-I characteristics of interconnection residual resistance, SMU
Pulse (A.04.00)
[Supported Analyzer]
B1505A
[Description]
Measures the residual resistance vs current characteristics of Interconnection. SMU pulse is used for current
output.
[Device Under Test]
Interconnection, 2 terminals
[Device Parameters]
YAxisResidualRMin: Y axis (ResidualR) minimum value
YAxisResidualRMax: Y axis (ResidualR) maximum value
[Test Parameters]
Memo: Memorandum
Port1: SMU connected to Port1 terminal, primary sweep current output
IStart: Sweep start current for Port1 terminal
IStop: Sweep stop current for Port1 terminal
ILinearStep: Sweep step current for Port1 terminal, effective if Scale=LINEAR
VLimit: Voltage compliance
PulsePeriodMode: Pluse period mode, AUTO or MANUAL
ManualPulsePeriod: Pulse period, effective if PulsePeriodMode=MANUAL
PulseWidth: Pulse width
Port2: GNDU connected to Port2 terminal
Scale: Scale of sweep, LINEAR, LOG10, LOG25, or LOG50
[Extended Test Parameters]
PulseBase: Pulse base current
PulseAvgCnt: Pulse averaging count
HoldTime: Hold time
MeasurementTime: Actural measurement time for a pulse period
[Measurement Parameters]
Measure voltage Vm
[User Function]
ResidualR: Residual resistance ResidualR=Vm/If
Ta: Temperature Ta=Temp
[X-Y Plot]
For Scale=LINEAR:
X axis: Force current If (LINEAR)
Y axis: Residual resistance ResidualR (LINEAR)
For Scale=LOG10, LOG25, or LOG50:
X axis: Force current If (LOG)
Y axis: Residual resistance ResidualR (LOG)
[List Display]
Agilent EasyEXPERT Application Library Reference, Edition 8
23-3