EasyManua.ls Logo

Agilent Technologies EasyEXPERT

Agilent Technologies EasyEXPERT
610 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24 MISCAP
24.1 BV: MISCAP Gate-Body breakdown voltage measurement (A.04.00)
[Supported Analyzer]
B1505A
[Description]
Measures MISCAP Gate current vs voltage characteristics and extracts the Gate-Body breakdown voltage.
[Device Under Test]
MISCAP, 2 terminals
[Device Parameters]
Polarity: Nch (SMUs force the specified value) or Pch (SMUs force the negative specified value).
Temp: Temperature
[Test Parameters]
Memo: Memorandum
IntegTime: Integration time
Gate: SMU connected to Gate terminal, primary sweep current output
IgStart: Sweep start current for Gate terminal
IgStop: Sweep stop current for Gate terminal
IgStep: Sweep step current for Gate terminal
Body: GNDU connected to Body terminal
Ileak@BV: Gate current to decide breakdown voltage BV
VgLimit: Gate voltage compliance
[Extended Test Parameters]
HoldTime: Hold time
DelayTime: Delay time
VgMinRange: Minimum range for gate voltage measurement
[Measurement Parameters]
Gate voltage Vgate
[User Function]
Ta: Temperature Ta=Temp
[Analysis Function]
BV=@MX (X coordinate of Marker)
[Auto Analysis]
Marker: Igate=Ileak@BV
[X-Y Plot]
X axis: Gate voltage Vgate (LINEAR)
Y axis: Gate current Igate (LINEAR)
[List Display]
Gate current Igate
Gate voltage Vgate
[Parameter Display Area]
Gate-Body breakdown voltage BV
Temperature Ta=Temp
Agilent EasyEXPERT Application Library Reference, Edition 8