24 MISCAP
24.3 Ileak-V: MISCAP I-V characteristics (A.04.00)
[Supported Analyzer]
B1505A
[Description]
Measures MISCAP I-V characteristics and extracts Gate-Body breakdown voltage.
[Device Under Test]
MISCAP, 2 terminals
[Device Parameters]
Polarity: Nch (SMUs force the specified value) or Pch (SMUs force the negative specified value).
Temp: Temperature
[Test Parameters]
Memo: Memorandum
IntegTime: Integration time
Gate: SMU connected to Gate terminal, primary sweep voltage output
VgStart: Sweep start voltage for Gate terminal
VgStop: Sweep stop voltage for Gate terminal
VgStep: Sweep step voltage for Gate terminal
Body: GNDU connected to Body terminal
Ileak@BV: Gate current to decide breakdown voltage BV
IgLimit: Gate current compliance
[Extended Test Parameters]
HoldTime: Hold time
DelayTime: Delay time
IgZero: Y axis (Ig) minimum value
IgMinRange: Minimum range for gate current measurement
[Measurement Parameters]
Gate current Igate
[User Function]
Ta: Temperature Ta=Temp
[Analysis Function]
BV=@MX (X coordinate of Marker)
[Auto Analysis]
Marker: Igate=Ileak@BV
[X-Y Plot]
X axis: Gate voltage Vgate (LINEAR)
Y1 axis: Gate current Igate (LINEAR)
Y2 axis: Gate current Igate (LOG)
[List Display]
Gate voltage Vgate
Gate current Igate
[Parameter Display Area]
Gate-Body breakdown voltage BV
Temperature Ta=Temp