25 PowerBJT
25.2 Ic-Vce(PowerBJT): Ic-Vce characteristics, SMU Pulse (A.05.00)
[Supported Analyzer]
B1505A
[Description]
Measures Ic-Vce characteristics. SMU pulse is used for the Collector-Emitter voltage output.
[Device Under Test]
Power BJT, 3 terminals
[Device Parameters]
Polarity: NPN (SMUs force the specified value) or PNP (SMUs force the negative specified value).
Temp: Temperature
[Test Parameters]
Memo: Memorandum
Collector: SMU connected to Collector terminal, primary sweep voltage output
VcStart: Sweep start voltage for Collector terminal
VcStop: Sweep stop voltage for Collector terminal
VcLinearStep: Sweep step voltage for Collector terminal
PulsePeriodMode: Pluse period mode, AUTO or MANUAL
ManualPulsePeriod: Pulse period, effective if PulsePeriodMode=MANUAL
VcPulseWidth: Pulse width of Collector voltage
IbPulseWidth: Pulse width of Base current
Base: SMU connection to Base terminal, secondary sweep current output
IbStart: Sweep start current for Base terminal
IbStop: Sweep stop current for Base terminal
VgStep: Sweep step current for Base terminal
Emitter: GNDU connected to Emitter terminal
IcLimit: Collector current compliance
PcLimit: Collector power compliance
[Extended Test Parameters]
VcPulseBase: Pulse base voltage for Collector terminal
IbPulseBase: Pulse base current for Base terminal
PulseAvgCnt: Pulse averaging count
VbLimit: Base voltage compliance
VcLimit: Collector voltage compliance (applicable for UHCU/HVMCU/UHVU)
HoldTime: Hold time
VcPulseDelayTime: Delay time of Collector voltage pulse
IbPulseDelayTime: Delay time of Base current pulse
MeasurementTime: Actural measurement time for a pulse period
[Measurement Parameters]
Collector voltage Vcollector
Collector current Icollector
Base voltage Vbase
Base current Ibase
[User Function]
Ta: Temperature Ta=Temp
[X-Y Plot]
Agilent EasyEXPERT Application Library Reference, Edition 8
25-4