27 PowerMOSFET, PMIC, SiC
27.12 Vds-Vgs: Vds-Vgs characteristics, Rds-Vgs characteristics, SMU Pulse
(A.05.00)
[Supported Analyzer]
B1505A
[Description]
Measures Vds-Vgs characteristics. SMU pulse is used for the Drain current output.
[Device Under Test]
Power MOSFET, 3 terminals
[Device Parameters]
Polarity: Nch (SMUs force the specified value) or Pch (SMUs force the negative specified value).
Temp: Temperature
[Test Parameters]
Memo: Memorandum
Gate: SMU connected to Gate terminal, primary sweep voltage output
VgStart: Sweep start voltage for Gate terminal
VgStop: Sweep stop voltage for Gate terminal
VgStep: Sweep step voltage for Gate terminal
PulsePeriodMode: Pluse period mode, AUTO or MANUAL
ManualPulsePeriod: Pulse period, effective if PulsePeriodMode=MANUAL
PulseWidth: Pulse width
Drain: SMU connected to Drain terminal, secondary sweep current output
IdStart: Sweep start current for Drain terminal
IdStep: Sweep current step for Drain terminal
IdPoint: Number of Drain current sweep steps
Source: GNDU connected to Source terminal
VdLimit: Drain voltage compliance
PdLimit: Drain power compliance
[Extended Test Parameters]
PulseBase: Pulse base current
PulseAvgCnt: Pulse averaging count
IgLimit: Gate current compliance
VdStop: Sweep stop Drain voltage (applicable to UHCU/HVMCU/UHVU)
HoldTime: Hold time
MeasurementTime: Actural measurement time for a pulse period
VdsScale: Graph scale option for Vds
VdsZero: Log scale minimum value for Vds
RdsScale: Graph scale option for Rds
RdsZero: Log scale minimum value for Rds
RdsMax: Scale maximum value for Rds
[Measurement Parameters]
Drain voltage Vds
Gate current Ig
Drain-Source resistance Rds
[User Function]
Ta: Temperature Ta=Temp
Agilent EasyEXPERT Application Library Reference, Edition 8