EasyManuals Logo
Home>Agilent Technologies>Water Pump>G5654A

Agilent Technologies G5654A User Manual

Agilent Technologies G5654A
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #101 background imageLoading...
Page #101 background image
Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 101
Error Information
6
General Error Messages
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status
signals. A LOW signal input on pin 6 of the Enhanced Remote Interface (ERI)
connector generates the error message.
Remote Timeout
Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one
minute of starting the analysis. If a not-ready condition is still present on the
remote line after one minute the error message is generated.
Probable cause Suggested actions
1
Leak detected in another module with a
CAN connection to the system.
Fix the leak in the external instrument before
restarting the module.
2
Leak detected in an external instrument
with a remote connection to the system.
Fix the leak in the external instrument before
restarting the module.
3
Shut-down in an external instrument with a
remote connection to the system.
Check external instruments for a shut-down
condition.
4
The degasser failed to generate sufficient
vacuum for solvent degassing.
Check the vacuum degasser for an error
condition. Refer to the Service Manual for the
degasser or the pump that has the degasser
built-in.
Probable cause Suggested actions
1
Not-ready condition in one of the
instruments connected to the remote line.
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up
correctly for analysis.
2
Defective remote cable.
Exchange the remote cable.
3
Defective components in the instrument
showing the not-ready condition.
Check the instrument for defects (refer to the
instrument’s documentation).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G5654A and is the answer not in the manual?

Agilent Technologies G5654A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG5654A
CategoryWater Pump
LanguageEnglish

Related product manuals