EasyManuals Logo

Agilent Technologies Infiniium DCA User Manual

Agilent Technologies Infiniium DCA
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #352 background imageLoading...
Page #352 background image
25-14
Waveform Commands
YREFerence?
Returned Format [:WAVeform:YRANge] <value><NL>
<value> A real number representing the Y-axis duration of the displayed waveform.
Example 10 OUTPUT 707;":SYSTEM:HEADER OFF”
20 OUTPUT 707;":WAVEFORM:YRANGE?"
YREFerence?
Query :WAVeform:YREFerence?
This query returns the level associated with the Y origin for the currently specified source. It
is at this level that the Y origin is defined. In this analyzer, the value is always zero.
Returned Format [:WAVeform:YREFerence] <integer_value><NL>
<integer_value> Always 0.
Example 10 OUTPUT 707;":SYSTEM:HEADER OFF”
20 OUTPUT 707;":WAVEFORM:YREFERENCE?"
See Also You can obtain the YReference value through the WAVeform:PREamble query.
YUNits?
Query :WAVeform:YUNits?
This query returns the Y-axis units of the currently selected waveform source. The currently
selected source may be a channel, function, waveform memory, TDR response, or color
grade/gray scale data.
Returned Format [:WAVeform:YUNits] {UNKNown | VOLT | OHM | SECond | REFLect | CONStant | AMP | WATT}<NL>
Example 10 DIM Unit$[50] !Dimension variable
20 OUTPUT 707;":WAVEFORM:YUNITS?"
30 ENTER 707;Unit$

Other manuals for Agilent Technologies Infiniium DCA

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infiniium DCA and is the answer not in the manual?

Agilent Technologies Infiniium DCA Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniium DCA
CategoryTest Equipment
LanguageEnglish

Related product manuals