OPERATION MANUAL OF DIGITAL LCR METER
OPERATION MANUAL OF DIGITAL LCR METEROPERATION MANUAL OF DIGITAL LCR METER
OPERATION MANUAL OF DIGITAL LCR METER
1
11
1
General
GeneralGeneral
General
This meter is a steady performance, battery-driven 3 1/2 digital LCR meter, which
has the figure of unit symbol displaying, accuracy reading and wide range
measurement. This meter has the function of measuring inductance, resistance,
capacitance test. In addition, its ideal for testing SMD type components. Plus,
the meter offers advanced features, such as the ability to perform precision
measurements. It is an ideal hand-held tool for labs, factories and
radio-technology.
Please read this operation manual carefully before your operation!
2
22
2
Characteristic
CharacteristicCharacteristic
Characteristic
2
22
2-
--
-1
11
1
General characteristic
General characteristicGeneral characteristic
General characteristic
DisplayLCD display, Max. 1999
Inductance measure0.1uH20H
Capacitance measure0.1pF2000uF
Resistance measure0.01Ω20MΩ
Zero adjustCapacitance under 20nF is manual ministrant zero adjust.
Over rangedisplay1
Low battery indication
symbol display
Sampling rate0-5s
Working temperature040
Relative humidity less than 80%
Meter size185mm×93mm×35mmlength×width×height
WeightApprox.290gincluding 9V battery
Powerone 9V battery (6F22 or equivalent
Standard accessoriestest leads, operation manual
2
22
2-
--
-2
22
2
Electric characteristic
Electric characteristicElectric characteristic
Electric characteristic
Accuracy ±(% of reading + number of digit) at 23 ± 5, relative
humidity<75% . One year guarantee since production date.
2-2-1Capacitance
Range Accuracy Resolution
200pF 0.1pF
2nF 1pF
20nF 10pF
200nF 100pF
2μF 1000pF
20μF 0.01μF
200μF
±(2.5%+5)
0.1μF
2000μF ±(3.0%+5) 1μF
2-2-2Inductance
Range Accuracy Resolution
200uH 0.1uH
2mH 1uH
20mH 10uH
200mH 100uH
2H
±(2.0%+5)
1mH
20H ±(4.0%+10) 10mH
2-2-3RESISTANCE
Range Accuracy Resolution
20Ω ±(1.0%+5) 0.01Ω
200Ω 0.1Ω
2kΩ 1Ω
20kΩ 10Ω
200kΩ
±(0.8%+2)
100Ω
20MΩ ±(2.0%+5) 10kΩ
3
33
3
Operation
OperationOperation
Operation
3
33
3-
--
-1
1 1
1
Panel descriptionSEE FIG.1
1LCD displaydisplay the measuring value, unit and
symbol;
2Power switchturn on/off the power;
3 Zero Adjust switch zero adjust switch before
measurement under range 20nF;
4Function knobselect range
5Input jacktest input COM;
6Battery case.
3
33
3-
--
-2
22
2
Matters need attention before measurement
Matters need attention before measurementMatters need attention before measurement
Matters need attention before measurement
When signal displays at the state of power on, should replace the battery
to ensure the accuracy;
Do not switch the function knob when measuring.
Capacitance measurement function is used to measure the capacitance value,
cant used to measure the figure of merit of reactive power. It perhaps gets
a wrong reading if measure the capacitance value of resistance; before measuring,
it should release the capacitor completely. Beware of polarity when connects
the polar capacitor. Do not connect the test COM with voltage source; otherwise
it will cause grievous damage. Do not connect the test leads short-circuit,
All manuals and user guides at all-guides.com
all-guides.com

Other manuals for AIDETEK DM4070

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AIDETEK DM4070 and is the answer not in the manual?

AIDETEK DM4070 Specifications

General IconGeneral
BrandAIDETEK
ModelDM4070
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals