15C0233-00 Rev A SSC4500DX 9/23/2020
Operation and Maintenance Manual
Sterile Storage Cabinet
SSC4500DX
99C0177-00-Standard unit
94C0078-00 Ceiling Duct Kit
91C0191-00 0-10VDC Retransmit Kit
91C0192-00 UV-C Light Kit Air Side
7000 Performance Drive
Syracuse, NY 13212
Phone (315) 452-7400
Fax (315) 452-7420
airinnovations.com

Questions and Answers

  • C
    Christine MannAug 11, 2025
    What to do if the Air innovations SSC4500DX display screen is blank?
    • N
      Noah HolmesAug 11, 2025
      If the display screen is blank, first, check if the unit has power by verifying the supply power. Next, ensure that the RJ45 cables are correctly connected at both ends. Finally, check the 24VDC circuit breaker; test for 24VDC power at the terminal blocks and reset the 24VDC circuit breaker if tripped.
  • J
    Joshua SchultzAug 18, 2025
    What to do if the Air innovations SSC4500DX Storage temperature is not holding the set point?
    • J
      Jacob BarnesAug 18, 2025
      If the temperature isn't holding, start by verifying that the set point is the correct value. Then, check if the heater circuit breaker has tripped and reset it if necessary. Also, make sure the door hasn't been open for too long; ensure it's closed and allow the unit to stabilize for one hour.
  • T
    Tyrone PooleAug 25, 2025
    Why does my Air innovations SSC4500DX Storage have low humidity?
    • P
      Phyllis DavidAug 25, 2025
      If you're experiencing low humidity, first check if there is water in the tank and add water if needed. Also, verify that the door hasn't been open for an extended time; ensure it is closed and allow the unit to stabilize for one hour. Finally, check if the humidifier circuit breaker has tripped and reset it if necessary.
  • G
    Gregory SerranoAug 29, 2025
    What to do if I get an Internal Pressure Alarm on my Air innovations SSC4500DX?
    • A
      Alexander LaneAug 29, 2025
      If you receive an internal pressure alarm, first verify that the door is closed and sealed, then allow the unit to stabilize for one hour. Also, check the external vent tube for any blockages and clear the pressure tube near the electrical connection if needed.
  • K
    Kendra LarsenSep 4, 2025
    How to fix a keyless lock that is not responding on Air Innovations Storage?
    • M
      mcneilerikaSep 4, 2025
      If the keyless lock isn't responding, it might be due to a wrong security code. Try using the master code to open and reset the keyless lock. Alternatively, the batteries in the lock might be expired; use the emergency open procedure and change the batteries.