What to do if there is insufficient gas flow rate in Air Liquide SGA2 Control Panel?
- AamillerAug 11, 2025
If you're experiencing an insufficient gas flow rate with your Air Liquide Control Panel, it could be due to several reasons. First, ensure that all valves are completely open, as a limited flow section through a valve can restrict the flow. If the cylinder is discharged or empty, replace it. If the cylinder valve isn't operating, replace the cylinder or bundle. In case the equipment is undersized, consult Air Liquide. Finally, if the equipment in output is not operational, replace it.