the
display
indicates
"SPAN"
and
the
span
concentration.
9.
Install
the
regulator
on
the
span
calibration
cylinder
and
connect
tub¬
ing
from
the
regulator
to
the
inlet
fit¬
ting
on
the
calibration
collar.
10.
Use
and
^
keys
to
adjust
the
concentration
displayed
on
the
IAQ50
to
match
the
concentration
on
the
span
gas
cylinder.
11.
Press
and
release
the
READ
key
to
take
a
span
gas
reading.
The
IAQ50
starts
a
60
second
countdown.
When
the
countdown
reaches
zero,
the
in¬
strument
returns
to
the
normal
meas¬
urement
mode.
12.
With
the
gas
still
connected,
observe
the
reading
on
the
display.
It
should
indicate
the
span
gas
concentration.
If
not,
repeat
the
calibration.
13.
If
the
displayed
reading
is
accurate,
remove
the
regulator
and
tubing.
14.
Turn
the
IAQ50
off
and
flip
DIP
switch
#7
to
OFF
(normal
mode).
The
calibra¬
tion
is
now
complete.