it.
When
finished,
turn
the
instrument
off
and
flip
DIP
switch
#7
to
"off."
On
the
Model
IAQ50,
press
and
hold
the
TEMP
or
HUMIDITY
key
to
enter
the
calibration
mode
for
that
parameter.
To
enter
the
calibration
mode
for
the
RTD,
make
sure
the
RTD
is
plugged
in.
Press
the
TEMP
key
until
"Prb"
shows
on
the
display,
then
press
and
hold
the
TEMP
The
display
will
begin
a
countdown
from
5
to
0.
Release
the
key
when
0
is
displayed.
The
top
line
of
the
display
will
show
the
adjustment
number
(factory
setting
is
0.0)
and
the
bottom
line
will
show
the
actual
reading
value.
Use
^
or
^
keys
to
change
the
adjustment
number
and
press
ENTER
to
accept.
If
necessary,
repeat
the
above
sequence
for
the
other
parameters.
When
finished,
turn
the
instrument
off
and
flip
DIP
switch
#7
to
"off."