To minimize the risk of transmission of blood-borne pathogens, the cleaning and disinfection procedure
should be performed as recommended in the instructions below.
The meter should be cleaned and disinfected after use on each patient. This blood glucose
monitoring system may only be used for testing multiple patients when standard precautions and the
manufacturers disinfection procedures are followed.
Cleaning and Disinfection
The cleaning procedure is needed to clean dirt, blood and other bodily fluids off the exterior of the
meter before performing the disinfection procedure.
The disinfection procedure is needed to prevent transmission of blood-borne pathogens.
Clorox
®
Germicidal Wipes, with 0.55% sodium hypochlorite as the active ingredient, have been tested
and approved for cleaning and disinfecting the Assure
®
Prism multi blood glucose meter. This disinfectant
wipe has been shown to be safe for use with this meter. Please read the Clorox
®
Germicidal Wipes
instructions before using on your meter.
Any disinfectant product with the EPA registration number of 67619-12 may be used on this device.
A list of Environmental Protection Agency (EPA) registered disinfectants effective against HIV, Hepatitis B,
and C virus can be found at the following website:
http://www.epa.gov/oppad001/list_d_hepatitisbhiv.pdf
Go to www.arkrayusa.com/diabetes-management/disinfect for the most up-to-date information
on approved cleaning and disinfecting procedures and materials for ARKRAY blood glucose
meters. If you have any questions, please contact our Technical Customer Service department at
800.818.8877,
24 hours a day, 7 days a week, USA and Canada
.
Name Clorox
®
Germicidal Wipes
Manufacturer
Clorox Professional Products Company®
800.537.1415
www.cloroxprofessional.com
EPA Registration Number 67619-12
Active Ingredients 0.55% Sodium Hypochlorite
Cleaning and Disinfecting your Assure® Prism multi Blood Glucose Monitoring System
Question and Answer IconNeed help?

Do you have a question about the Akray Assure Prism Multi and is the answer not in the manual?

Akray Assure Prism Multi Specifications

General IconGeneral
Sample TypeCapillary whole blood
Test Time5 seconds
Operating Humidity10 - 90% RH
DisplayLCD
Measurement Range20-600 mg/dL
Battery Type2 AAA batteries
Operating Temperature10 - 40°C (50 - 104°F)
Memory Capacity500 tests

Summary

Cleaning and Disinfecting the Assure Prism Multi System

Purpose and Approved Disinfectant

Explains the necessity of cleaning/disinfection and identifies the approved Clorox® Germicidal Wipes.

Cleaning and Disinfecting Procedures

Cleaning Process Steps

Details the steps for cleaning the meter exterior using disposable wipes to remove fluids.

Disinfecting Process Steps

Outlines the steps for disinfecting the meter exterior with a new wipe to prevent pathogen transmission.

Deterioration Signs and Support

Lists signs of meter deterioration after cleaning and contact information for customer service.