19.1 General maintenance instructions
19 Maintenance
The service life of the products (incl. accessories) described in these instructions for use is of course
dependent on the type of use.
In order to ensure safe operation, the product in question must be visually inspected
and functionally checked at regular intervals by appropriate qualified personnel, at least
once a year, every time the product is returned to use and after every repair, in
accordance with the maintenance schedule. Shorter inspection cycles may be necessary
if the product is used more frequently than normal.
EN ISO 10535 specifies the following in Annex B:
“Periodic inspection should be performed by a suitable and properly qualified person who is
familiar with the design, application and maintenance of the hoist.
If the product is not regularly and properly serviced, safe use is no longer guaranteed. Wear, damage
as well as loosening of connecting elements can therefore not be detected.
If any checks indicate defects/damage, immediately disconnect the product from the
mains supply and do not operate it again. Press the emergency stop button and remove
the battery pack (see chapter on Operation, section Emergency stop button
and
section Battery pack
). Mark the product clearly as “out of order” and inform your
authorised dealer immediately.
Unplug the product from the mains supply and press the emergency stop button
before the visual inspection to prevent danger, e.g. from damaged insulation
on power supply lines. If you do not identify any damage during your visual
inspection, unlock the emergency stop button for the function test.
Regularly and at short intervals check the power supply line for mechanical
damage (e.g. monthly and after any mechanical strains that occur).
Removed, defective/damaged electrical components may only be checked and
evaluated by a qualified electrician or aks GmbH.
Do not carry out any repairs on the product that would alter the product characteristics.
If this instruction is not observed, a safe power supply can no longer be guaranteed.
Furthermore, there is an increase to the risk that is always present when lifting/
uprighting and transferring people. In addition, any warranty claims and product liability
are excluded.
Maintenance may only be carried out on empty patient hoists.
Defective/damaged electrical components must be replaced by suitable qualified
personnel. Electrical components must not be opened and must be replaced as
complete units.
Exception: Replacement of the batteries (battery pack) by suitable trained and qualified
personnel.