Screen 1
- Limit Setup To go to the other menu
- SpO2 Hi Set hi levels of oxygen
percentage
- SpO2 Lo Set low levels of oxygen
percentage
- PR Hi Set high levels of pulse
rate
- PR Lo Set low levels of pulse rate
- +/- Switch between going
higher or lower
- Exit To exit the menu
Screen 2
- Remind Setup To go to the other menu
- Sound reminder To switch the alarm on or o󰀨
- Beep To switch the pulse rate
alarm on or o󰀨
- Demo to play a demo (power o󰀨 to
start again without the demo)
- Restore To restore back to factory
settings
- Brightness To set the display intensity
- Exit To exit the menu
Maintenance:
Cleaning and storage:
The device is non-sterile and not intended to be
sterilized.
Clean the black rubber nger sleeve before and
after using with medicinal alcohol (> 95%).
The exterior part of the tester can be cleaned
using a damp cloth. The ACS-31 is not water-
proof, so DO NOT immerse in water.
Keep the pulse Oximeter out of the reach of
children, and preferably inside the included
case in a dry place.
See the technical specications regarding
storage conditions.
Batteries:
The batteries can power the unit for over 20
hours continuously.
If the device will not be used for a long period
(> 2 months), remove the batteries. Doing so
prevent any possible damage from leaking
batteries.
Replace the batteries when the battery indi-
cator shows that the batteries are running low
(display: or ).
Always place two identical batteries, type
alkaline, size AAA.
Rechargeable batteries should not be used
due to their lower voltage.
Environment:
Should the device need to be discarded,
hand it over to your supplier. They
will dispose of it in an environmen-
tally friendly manner.
Empty batteries can be returned to the sup-
plier or disposed of in an environmentally-
friendly manner according to your local regu-
lations. Do not dispose of used batteries with
your normal household waste.
Technical:
24 months Warranty ● ACS-31 does not con-
tain any service components/accessories ●
maintenance only through the service depart-
ment of Alecto.
Calibration:
The ACS-31 does not need to be calibrated.
Declaration of Conformity:
http://DOC.hesdo.com/ACS-31-DOC.pdf
Specications:
Display: Colour, type OLED
Pulse rate:
Range: 30bpm ~ 250bpm
Resolution: 1bpm
Accuracy: 1bpm
Strength: bar chart (bpm = beats per minute)
SpO2:
Range: 35 -100%
Resolution: 1%
Accuracy: 90% -100%: ±2%
70% - 90%: ±3%
< 70%: unspecied
Perfusion Index:
Range: 0-20%
Resolution: 0,1%
Accuracy: 0% - 1%: 0,1%
1% - 20%: 1%
Batteries: 2x AAA, 1.5V, Alkaline
Consumption: 30mA (nominal)
Battery life: >20 hours continuous use
Dimensions, weight:
62,8 x 38 x 29mm / 51 grams (incl. batt)
Surroundings:
Operating temperature: 5°C ~ 40°C
Storage temperature: -10°C ~ 50°C
Operating humidity: 15% ~ 80%
Storage Humidity: 10% ~ 93%
: Shanghai International Holding Corp.
GmbH (Europe).
Ei󰀨estrasse 80, 20537, Hamburg, Germany.
: Shenzen AEON Technology Co. LTD.
RM6H02, Block 27-29, Tianxia IC Industrial Park
Majialong, No. 133 of Yiyuan Road, Nantou
Street, Nanshan District, Shenzen, China.