FAILURE TO FOLLOW INSTRUCTIONS AND/OR
IMPROPER USE MAY RESULT IN SCALDING,
BODILY INJURIES OR EXPLOSION.
When using the pressure steam sterilizer, basic
safety precautions should always be followed:
1. Read and understand instruction manual before
operating unit.
2. Do not touch hot surfaces. Use handles and pot
holders.
3. Close supervision is necessary when the
sterilizer is used near children.
4. Extreme caution must be used when moving a
sterilizer containing hot liquids.
5. Do not use the sterilizer for other than intended
use.
6. Always check the pressure release devices for
clogging before use.
7. This sterilizer operates under pressure.
Improper use may result in scalding injury.
Make certain unit is properly closed before
operating. Read Operating Instructions.
8. Never loosen wing nuts until the steam pressure
gauge registers zero and you have allowed any
remaining pressure to escape by opening the
control valve (lever in the vertical position).
9. Do not open the sterilizer until the unit has
cooled and internal pressure has been reduced.
Gauge should read zero at this time. Read
Operating Instructions.
10.Never use the sterilizer for cooking or
processing food.
11.Never place oil in or on this sterilizer.
12. Do not subject your sterilizer to sudden extreme
temperature changes, as this will cause
expansion or contraction which can crack a cast
aluminum utensil. Do not move a sterilizer from a
cold storage area directly onto a hot flame or
element. Do not add cold water to a sterilizer
which has boiled dry and is still hot. Do not cool
the sterilizer suddenly by pouring cold water on it
or wrapping cold wet towels around it.
13. Always operate sterilizers on surfaces that will
not be damaged by heat. We recommend the use
of our support base. See page 5, item 6.
14. As in all clinical laboratory settings, wear safety
glasses when attending to your sterilizer.
SAVE THESE INSTRUCTIONS
CAUTION! READ THESE IMPORTANT SAFEGUARDS!
25X shown with optional support base, Part No. 2180
OPERATING INSTRUCTIONS
Model 25X Electric Pressure Steam Sterilizer
REV 4/09
Question and Answer IconNeed help?

Do you have a question about the All American 25X-120 and is the answer not in the manual?

Summary

Safety and Environment Guidelines

General Safety Precautions

Essential safety rules and warnings for operating the sterilizer to prevent injury or explosion.

Operating Environment Guidelines

Conditions for optimal and safe operation, including temperature and humidity limits.

Altitude Adjustment for Pressure

How elevation affects required steam pressure settings for accurate sterilization.

Operating the Sterilizer

Lubricate Metal-to-Metal Seal

Applying lubricant to the metal-to-metal seal for proper function and to prevent sticking.

Exhausting the Sterilizer

Vigorously releasing steam to remove trapped air for at least five minutes.

Heat Control Knob Operation

Adjusting the heat control knob for desired temperature and pressure.

Sterilization Period Timing

Timing the sterilization cycle for a minimum of 35 minutes after reaching pressure.

Post-Sterilization Cooling and Release

Safely cooling the unit and releasing internal pressure before opening.

Thermostat Calibration Instructions

Instructions for calibrating the thermostat for accurate temperature control.

Maintenance and Component Details

Metal-to-Metal Seal Maintenance

Periodic lubrication and cleaning of the metal-to-metal seal to prevent damage.

Pressure Gauge Check

Ensuring the pressure gauge opening is clear and functional.

Control Valve Cleaning

Periodic cleaning of the control valve for optimal performance and longevity.

Air Exhaust Tube Check

Ensuring the air exhaust tube is clear and unobstructed for proper venting.

Excess Pressure Relief Valve Maintenance

Routine replacement and checking of the excess pressure relief valve.

Overpressure Plug Function and Maintenance

Details on the overpressure plug's role, replacement, and cleaning for safety.

Excess Pressure Relief Valve Operation

Explanation of how the excess pressure relief valve functions under pressure.

Technical Information and Specifications

Sterilization Principles and Failure Causes

Explains steam's effectiveness and air trapping as a cause of sterilization failure.

Pressure Gauge Accuracy Details

Details on gauge accuracy standards and specifications.

Tips for Wet Bags or Packs

Suggestions for preventing wet bags or packs during sterilization.

Model 25 X Specifications

Detailed technical specifications for the Model 25X sterilizer.

Other Electric Model Specifications

Specifications for other electric sterilizer models offered by the manufacturer.

Electrical Wiring Pictorial Diagram

Visual representation of the sterilizer's electrical connections.

Electrical Wiring Schematic Diagram

Schematic diagram detailing the sterilizer's electrical circuit.

Warranty and Service Information

Limited Warranty Information

Details of the manufacturer's warranty for the sterilizer.

Return and Service Instructions

Procedures for returning the unit for service or repairs.

All American 25X-120 Specifications

General IconGeneral
CategoryLaboratory Equipment
Power SourceElectric
TypeAutoclave/Sterilizer
Power Requirements120V, 60Hz
MaterialAluminum