Ventilator Connections
Oxygen Connection (if O
2
is available)
A 40-80 psi compressed oxygen source can be connected to the MCV100
to deliver 100% oxygen to the patient. 100% or 60% oxygen can be
delivered to the patient with the MCV100-B.
Connect a compressed oxygen source by attaching one end of an oxygen
hose to the oxygen DISS inlet shown above, and then connecting the other
end to an oxygen regulator, or hospital or ambulance oxygen outlet.
Ensure that the oxygen source is pressurized and delivering oxygen.
Ensure that the unit is set to 100% O
2
on the MCV100, or 100% or 60% O
2
on the MCV100-B
MCV100 -B Source Gas Selection
8