Microwash Q-Rinse System

Questions and Answers

  • P
    Patrick StanleyAug 15, 2025
    What to do if my Altomed Medical Equipment unit fails to operate?
    • S
      Samuel SkinnerAug 15, 2025
      If the Altomed Medical Equipment fails to operate, first, ensure the unit is properly plugged into a functioning power outlet, referring to figure 5a. If plugged in, verify the power outlet and switch (figure 5e), and check the green pressure line connection at the back of the unit (figure 5b). As a last resort, let the unit sit unplugged for 10 minutes with the green pressure line connected (figure 5b). If these steps do not resolve the issue, the internal parts may have failed, and the machine should be sent for repair.
  • D
    Dennis EvansAug 18, 2025
    Why is there poor pressure of both fluid and air in my Altomed Microwash Q-Rinse?
    • B
      Barry SchmidtAug 18, 2025
      If you're experiencing poor pressure of both fluid and air in your Altomed Medical Equipment, start by checking the green pressure line for a secure connection at the back of the unit (figure 5b). Also, inspect the line for any cracks or leaks. Make sure that the cap or connector gaskets aren't damaged or missing by checking inside the cap for the gasket and inspecting the tubing connectors for the presence and correct placement of black gaskets. If the issue persists, the pump unit may have failed, requiring the machine to be sent for repair.
  • M
    Maria GomezAug 21, 2025
    How to fix no or poor fluid flow in Altomed Microwash Q-Rinse Medical Equipment?
    • A
      Ashley KaneAug 22, 2025
      If you experience no or poor fluid flow in your Altomed Medical Equipment, begin by checking the fluid level in the bottle and ensuring the bottle cap is tight. Then check the green pressure line for a secure connection at the back of the system (figure 5b). Verify that the white gasket is present inside the cap and that the black gaskets on the tubing connectors are present and correct. If the fluid valve is not opening, do not use the machine and send it for repair.
  • R
    Ross ParkAug 25, 2025
    What to do if there is no flow of air in my Altomed Medical Equipment?
    • J
      Julie KleinAug 26, 2025
      If there is no air flow in your Altomed Medical Equipment, check the tightness of the bottle cap and ensure the green pressure line is securely connected at the back of the system (figure 5b). Also, verify that the white gasket is present inside the cap and that the black gaskets on the tubing connectors are present and correct. If the air valve is not opening, do not use the machine and send it for repair.
  • M
    Matthew LiAug 28, 2025
    What to do if my Altomed Microwash Q-Rinse Medical Equipment unit fails to switch modes?
    • T
      Thomas MahoneyAug 28, 2025
      If your Altomed Medical Equipment fails to switch modes, first, press the STOP button to reset the unit, in case the cycle wasn't completed from the last sequence. If this doesn't work, the cycle switches may have failed, and the machine should be sent for repair.
  • B
    brianaschmidtSep 1, 2025
    Why doesn't the foot pedal operate on my Altomed Medical Equipment?
    • D
      Donald SnyderSep 1, 2025
      If the foot pedal on your Altomed Medical Equipment does not operate, ensure it is properly plugged in (figure 5d).
  • H
    Henry GreenSep 4, 2025
    Why does my Altomed Medical Equipment start in the wrong mode?
    • A
      Albert Austin MDSep 4, 2025
      If your Altomed Medical Equipment starts in the wrong mode, press STOP to reset the unit. This will resolve the problem if the cycle was not completed from the last sequence.
  • G
    gardnerlisaSep 7, 2025
    What to do if fluid is backing up the interface tubing in Altomed Microwash Q-Rinse Medical Equipment?
    • M
      maddoxjamieSep 7, 2025
      If fluid is backing up the interface tubing in your Altomed Medical Equipment, control the flow of fluid by switching to the manual mode, as the incorrect cycle may be in use.
  • J
    Jessica ContrerasSep 12, 2025
    Why don't the tubing connectors fit inside the cap of my Altomed Medical Equipment anymore?
    • O
      Olivia HamptonSep 12, 2025
      If the tubing connectors no longer fit inside the cap of your Altomed Medical Equipment, replace the cap assembly. This is the suggested solution if the tubing has been disconnected from the bottle during sterilization.

Summary

Warning, Cautions and Symbols

Special Instructions; Notes (EMC)

Details EMC testing to EN 60601-1-2:2015 and special precautions for use.

Set up and Operation

Initial Setup Procedures

Guides the user through unpacking, pre-use checks, connections, and powering on.

Automated Cycle Operation

Details the steps for performing an automated rinse cycle using the system.

Manual Cycle Operation

Explains how to operate the system using the manual cycle mode for controlled rinsing.

Troubleshooting

Maintenance

System Specifications

Electromagnetic Compatibility (EMC) and Safety Notes

Customer and Technical Support

Altomed Microwash Q-Rinse Specifications

General IconGeneral
BrandAltomed
ModelMicrowash Q-Rinse
CategoryMedical Equipment
LanguageEnglish