5.29 Second calibration scaling point
Display: CAL2
Range: Any display value
Default Value: n/a
Second scaling point for 2 point calibration scaling - See “Calibration” chapter, section 6.1.
5.30 Calibration offset
Display: CAL OFSt
Range: Any display value
Default Value: n/a
Calibration offset - See section 6.3.
5.31 Zero range
Display: ZERO RNGE
Range: Any display value or OFF
Default Value: OFF
Zero range limit value - see section 6.4.
5.32 Zero reference point for ZERO RNGE operation
Display: CAL ZERO
Range: n/a
Default Value: n/a
Zero point calibration for ZERO RNGE function - see section 6.5.
5.33 4mA input scale
Display: USER En4
Range: Any display value
Default Value: n/a
4mA input scale value, use only as an alternative to CAL1 and CAL2 calibration - See “Calibra-
tion” chapter, section 6.2.
28 of 56 PM4IVMAN-2.4-0