Display shows “- - - -” - this message indicates that the input signal is higher than the range
selected. e.g. for an input set for 0-1V the “- - - -” message will be seen if the input signal goes
much higher than 2V.
Display shows -or- - this message indicates either that the number is too big to display e.g.
above 9999 on a 4 digit display or that the dISP function has been set to -or- and either the
Lo dISP or HIGH dISP limits have been exceeded.
Display value flashes on and off - this indicates that the dISP function has been set to FLSH
and either the Lo dISP or HIGH dISP limits have been exceeded.
Display shows NO ACC - this indicates that the ACCS function has been set to NONE or the
R.INP function has been set to no.Ac blocking entry to FUNC mode. Enter functions via CAL
mode to gain entry to functions and if required change the ACCS or R.INP function setting.
Display shows NO SPAC - this indicates that the R.INP function has been set to SP.Ac blocking
entry to alarm relay functions. Enter functions via CAL mode to gain entry to functions and if
required change the R.INP function setting.