Positions the Artex®CR on the table top with the front raised.
Releases the centric by lifting the pivoting lever, suspending condyle ball fixation.
Fixes the Artex top section in the central position by moving semi-axes outward.
Houses individually adjustable horizontal condyle path inclination, Bennett angle, ISS, protrusion, retrusion.
Adjusts the condyle path inclination angle from -20° to +60° using an Allen key.
Sets the Bennett angle between -5° to +30° using a locking screw and rotary knob.
Adjusts ISS individually and infinitely up to 1.5 mm per side using a locking screw.
Guides the top section protrusively, adjustable up to 6 mm by rotating a knurled screw.
Guides the top section retrusively by rotating a knurled screw, adjustable via a retrusion insert.
Describes the condyle ball's centered position and movement within the condylar box.
Sets distraction value (0-3 mm) by installing a loose screw from the frontal.
Lifts frame top for compression reading and positions models into the articulator.
Transfer of maxillary cranium/axis using facebow, referencing arbitrary pin and axis mean point.
Details protrusion track angle (40°) and lateral movement variation based on Bennett angle.
Emphasizes careful handling of precision instruments and lists cleaning agent precautions.
| Category | Medical Equipment |
|---|---|
| Manufacturer | Amann Girrbach |
| Model | Artex CR |
| Type | Articulator |
| Technology | Mechanical |
| Intercondylar distance | 110 mm |
| Bennett angle | Adjustable |
| Bilateral freedom of movement | Yes |
| Centric latch | Yes |
| Application | Dental |
| Compatibility | Dental Models |
| Dimensions | Varies by model |
| Weight | Varies by model |