Ampcontrol Pty Ltd ABN 28 000 915 542
GG2 COMMUNICATIONS USER MANUAL
GG2B032
Version 1 JULY/2020
Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version
APPROVED FOR EXTERNAL DISTRIBUTION
PROPERTY OF AMPCONTROL PTY LTD
NOT TO
BE REPRODUCED IN PART
Block 6
Only available via a Read/Write S/N. Not visible via regular rollcall page on the Controller.
Must be performed by
NATA certified 3rd
parties. E.g. CMTS, Cal
Pacific etc.
Last Linearity test,
number of test points
Bits 2 - 0 : 0 to 6 points
5,2
Post-box options :
0 : Linearity test
timestamp high-word
1 : Telemetry test start
timestamp high-word
2 : t90 test timestamp
high-word
3 : RTC current high-
word
Unix style timestamp,
using the Ampcontrol
epoch -
01-01-2000 00:00.
Big endian
Note, it will take the
GG2 at least 1.5 second
to change the post-box
value before it can be
read out. (this after the
write s/n is complete)
R
5,3
Post-box options :
0 : Linearity test
timestamp low-word
1 : Telemetry test start
timestamp low-word
2 : t90 test timestamp
low-word
3 : RTC current low-word
Block 6
Description
Value
R/W
Application Notes
6,0
Future use
Future use
W
6,1
Detector Serial number
high word
Detector SyteLine serial
number (10 digit)
32 bits (for calibration
tracking)
R
6,2
Detector Serial number
low word
6,3
Future use
MSB : Future use
R
Detector Software
Version
LSB : 8 bit version
number (unsigned)